特级婬片A片AAA毛片AA|97无码免费人妻超级碰碰夜夜|亚洲无码中文|亚洲成a人片在线观看无码专区
Ultrasonic measurement of optic nerve sheath diameter: a non-invasive surrogate approach for dynamic, real-time evaluation of intracranial pressure

摘自:Br J Ophthalmol: first published as 10.1136/bjophthalmol-2018-312934 on 25 October 2018.

作者:Li-min Chen, Li-juan Wang, Yang Hu, Xiao-han Jiang, Yu-zhi Wang, Ying-qi Xing

Abstract
The current study aimed to identify whether ultrasonographic measurements of optic nerve sheath diameter (ONSD) could dynamically and sensitively evaluate real-time intracranial pressure (ICP). ONSD measurements were performed approximately 5 min prior to and after a lumbar puncture (LP). A total of 84 patients (mean±SD age, 43.5±14.7 years; 41 (49%) men; 18 patients with elevated ICP) were included in the study. The Spearman correlation coefficients between the two observers were 0.779 and 0.703 in the transverse section and 0.751 and 0.788 in the vertical section for the left and right eyes, respectively. The median (IQR) change in ONSD (ΔONSD) and change in ICP (ΔICP) were 0.11 (0.05–0.21) mm and 30 (20–40) mmH2O, respectively, for all participants. With a reduction in cerebrospinal fluid pressure, 80 subjects (95%) showed an immediate drop in ONSD; the median (IQR) decreased from 4.13 (4.02–4.38) mm to 4.02 (3.90–4.23) mm (p<0.001). Significant correlations were found between ONSD and ICP before LPs (r=0.482, p<0.01)
and between ΔONSD and ΔICP (r=0.451, p<0.01). Ultrasonic measurement of ONSD can reflect the relative real-time changes in ICP.

Ultrasonic measurement.pdf

特级婬片A片AAA毛片AA|97无码免费人妻超级碰碰夜夜|亚洲无码中文|亚洲成a人片在线观看无码专区